59
Match Measurement Uncertainty
Figure 6 Match Measurement Uncertainty
Measurement Uncert.
burst width > 200 s
1 s < b.w. < 200 s
0.5 s < burst width < 1 s
burst width < 0.5 s
5012B ± 7% of reading, +0.20 W
5016B ± 7% of reading, + 0.05 W
5017B ± 7% of reading, + 0.70 W
5018B ± 7% of reading, + 0.05 W
5019B ± 7% of reading, +0.13 W
5012B ± 10% of reading, + 0.40 W
5016B ± 10% of reading, + 0.10 W
5017B ± 10% of reading, + 1.40 W
5018B ± 10% of reading, + 0.10 W
5019B ± 10% of reading, + 0.26 W
5012B ± 15% of reading, + 0.40 W
5016B ± 15% of reading, + 0.10 W
5017B ± 15% of reading, + 1.40 W
5018B ± 15% of reading, + 0.10 W
5019B ± 15% of reading, + 0.26 W
5012B ± 20% of reading, + 0.40 W
5016B ± 20% of reading, + 0.10 W
5017B ± 20% of reading, + 1.40 W
5018B ± 20% of reading, + 0.10 W
5019B ± 20% of reading, + 0.26 W
*. Max. power depends on frequency and system
VSWR. See Figure 7 on page 60
Above 35 °C or below 15 °C add 3%
For D < 0.1 add 0.1 W
For period > 0.1s add (1.5% + 0.15 W)